Abstract

\textit{Topmetal-II−{II}^-} is a highly pixelated direct charge sensor that contains a 72×{\times}72 pixel array of 83μμm pitch size. The key feature of \textit{Topmetal-II−{II}^-} is that it can directly collect charges via metal nodes of each pixel to form two-dimensional images of charge cloud distributions. \textit{Topmetal-II−{II}^-} was proved to measure charged particles without amplification at room temperature. To measure its performance at cryogenic temperature, a \textit{Topmetal-II−{II}^-} sensor is embedded into a liquid nitrogen dewar. The results presented in this paper show that \textit{Topmetal-II−{II}^-} can also operate well at this low temperature with a noise (ENC) of 12 e−^- lower than that at room temperature (13 e−^-). From the noise perspective, \textit{Topmetal-II−{II}^-} is a promising candidate for the next generation readout of liquid argon and xenon Time Projection Chamber (TPC) used in experiments searching for neutrinoless double beta decay and dark matter.