Abstract
A feasible and convenient method has been proposed to suppress higher-harmonics for varied-line-spacing (VLS) plane grating monochromator in soft x-ray region. Related calculations and experiments demonstrate that decreasing the included angle slightly by changing the parameter of exit arm length can significantly improve light purity. This method is suitable and has been used for experiments of detector calibration in beamline 4B7B at Beijing Synchrotron Radiation Facility (BSRF).