Abstract
We demonstrate a single-shot terahertz spectrometer consisting of a modified Mach-Zehnder interferometer and a microbolometer focal plane array. The spectrometer is simple to use and can measure terahertz field autocorrelations and spectral power with no moving parts and no ultrashort-pulsed laser. It can effectively detect radiation at 1040 THz when tested with a thermal source. It can be also used to measure the complex refractive index of a sample material. In principle, it can characterize both laser-based and non-laser-based terahertz sources and potentially cover 110 THz with specially-designed terahertz microbolometers.