Abstract
Polarization detection of X-rays is a non-negligible topic to astrophysical observation. Many polarization detection methods have been well developed for X-rays in the energy range below 10~keV, while the detection at 10-30~keV is rarely discussed. This paper presents a simulation study of a Xe-based gas pixel detector, which can achieve the polarization detection of X-rays at 10-30~keV. To verify the emission angle distribution of photoelectrons, different electromagnetic models in Geant4 were investigated. After a necessary modification by considering the missing factor when sampling the emission angle, a good agreement can be achieved. Moreover, the detection capability of 20~keV polarized photons was discussed and the modulation factor could be 43\%.
Connections
Explore this paper’s authors, topics and related work.
Building this map preview
BZPEER is loading the nearby papers, people, topics and institutions for this page.
Reviews 0
Write a review
No reviews yet.
Discussion 0
Add a comment
No comments yet.