Optimization of the X-ray incidence angle in photoelectron spectrometers

preprint2012arXivOpen access

Abstract

Interplay of the angle dependent X-ray reflectivity and absorption with the photoelectron attenuation length in the photoelectron emission process determine the optimal X-ray incidence angle which maximizes the photoelectron signal. Calculations in a wide VUV through hard-X-ray energy range show that the optimal angle goes with energy progressively more grazing from a few tens of degrees at 50 eV to about one degree at 3.5 keV accompanied by the intensity gain increasing up to a few tens of times as long as the X-ray footprint on the sample stays within the analyzer field of view. This trend is fairly material independent. The obtained results bear immediate implications for design of the (synchrotron based) photoelectron spectrometers.

Explore this paper’s authors, topics and related work.

Building this map preview

BZPEER is loading the nearby papers, people, topics and institutions for this page.

Reviews 0

Write a reviewWrite

No reviews yet.

Discussion 0

Add a commentWrite

No comments yet.