NaI (Tl) Calorimeter Calibration and Simulation for Coulomb Sum Rule Experiment in Hall-A at Jefferson Lab

preprint2010arXivOpen access

Abstract

A precision measurment of inclusive electron scattering cross sections was carried out at Jefferson Lab in the quasi-elastic region for 4{^4}He, 12{^{12}}C, 56{^{56}}Fe and 208{^{208}}Pb targets. Longitudinal (RLR_{L}) and transverse (RTR_{T}) response functions of nucleon were extracted in the momentum transfer range 0.55 GeV/c≤\le∣q∣|q|≤\le1.0 GeV/c. To achieve the above goal, a NaI (Tl) calorimeter was used to distinguish good electrons from background including pions and low energy electrons rescattered from walls of the spectrometer magnets. Due to a large set of kinematics and changes in HV settings, a number of calibrations were performed for the NaI (Tl) detector. Corrections for a few blocks of NaI (Tl) with bad or no signal were applied. The resolution of NaI (Tl) detector after calibration reached δEE≈3%\frac{δE}{\sqrt{E}} \approx 3\% at E=1 GeV. The performance of NaI (Tl) detector was compared with a simulation.