Abstract

We have developed a position response calibration method for a micro-channel plate (MCP) detector with a delay-line anode position readout scheme. Using an {\em in situ} calibration mask, an accuracy of 8~μμm and a resolution of 85~μμm (FWHM) have been achieved for MeV-scale αα particles and ions with energies of ∼\sim10~keV. At this level of accuracy, the difference between the MCP position responses to high-energy αα particles and low-energy ions is significant. The improved performance of the MCP detector can find applications in many fields of AMO and nuclear physics. In our case, it helps reducing systematic uncertainties in a high-precision nuclear ββ-decay experiment.