The β−β^- based radio-guided surgery overcomes the corresponding γγ technique in case the background from healthy tissues is relevant. It can be used only in case a radio-tracer marked with 90^{90}Y is available since the current probe prototype was optimized for the emission spectrum of this radio-nuclide. Here we study, with a set of laboratory tests and simulations, the prototype capability in case a different radio-nuclide is chosen among those used in nuclear medicine. As a result we estimate the probe efficiency on electrons and photons as a function of energy and we evaluate the feasibility of a radio-guided surgery exploiting the selected radio-nuclides. We conclude that requiring a 0.1~ml residue to be detected within 1~s by administering 3~MBq/Kg of radio-isotope, the current probe prototype would yield a significant signal in a vast range of values of SUV and TNR in case 31^{31}Si,32^{32}P, 97^{97}Zr, and 188^{188}Re are used. Conversely, a tuning of the detector would be needed to efficiency use 83^{83}Br, 133^{133}I, and 153^{153}Sm, although they could already be used in case of high SUV or TNR values. Finally, 18^{18}F,67^{67}Cu, 131^{131}I, and 177^{177}Lu are not useable for radio-guided surgery with the current probe design.