Abstract

In this work, we investigated the ability of a high-purity germanium detector connected to a trapezoid-filter-based data acquisition system to reliably record signals in spite of high sample activities. By activating multiple Na2_{2}CO3_{3} samples with different Na content, we were able to deduce efficiency, resolution and dead time of the system used as a function of the sample activity. Based on the results, we were able to find a setting which allows measurements of event rates up to 35~kHz per readout channel with an energy resolution of 0.3\% at the 2754 keV 24^{24}Na line.