Abstract
An unparalleled resolution is reported with an inelastic x-ray scattering instrument at the Cu K-edge. Based on a segmented concave analyzer, featuring single crystal quartz (SiO_{2}) pixels, the spectrometer delivers a resolution near 25 meV (FWHM) at 8981 eV. Besides the quartz analyzer, the performance of the spectrometer relies on a four-bounce Si(553) high-resolution monochromator and focusing Kirkpatrick-Baez optics. The measured resolution agrees with the ray tracing simulation of an ideal spectrometer. We demonstrated the performance of the spectrometer by reproducing the phonon dispersion curve of a beryllium (Be) single crystal.