Abstract
Pencil x-ray beam imaging provides superior spatial resolution than other imaging geometries like sheet beam and cone beam geometries due to the illumination of a line instead of an area or volume. However, the pencil beam geometry suffers from long scan times and concerns over dose discourage laboratory use of pencil beam x-ray sources. Molecular imaging techniques like XLCT imaging benefit most from pencil beam imaging to accurately localize the distribution of contrast agents embedded in a small animal object. To investigate the dose deposited by pencil beam x-ray imaging in XLCT, dose estimations from one angular projection scan by three different x-ray source energies were performed on a small animal object composed of water, bone, and blood with a Monte Carlo simulation platform, GATE (Geant4 Application for Tomographic Emission). Our results indicate that, with an adequate x-ray benchtop source with high brilliance and quasi-monochromatic properties like the Sigray source, the dose concerns can be reduced. With the Sigray source, the bone marrow was estimated to have a radiation dose of 30 mGy for a typical XLCT imaging, in which we have 6 angular projections, 100 micrometer scan step size, and 10^6 x-ray photons per linear scan.