Abstract

HgCdTe detector arrays with a cutoff wavelength of ~10 μμm intended for the NEOCam space mission were subjected to proton beam irradiation at the University of California Davis Crocker Nuclear Laboratory. Three arrays were tested - one with 800 μμm substrate intact, one with 30 μμm substrate, and one completely substrate-removed. The CdZnTe substrate, on which the HgCdTe detector is grown, has been shown to produce luminescence in shorter wave HgCdTe arrays that causes elevated signal in non-hit pixels when subjected to proton irradiation. This testing was conducted to ascertain whether or not full substrate removal is necessary. At the dark level of the dewar, we detect no luminescence in non-hit pixels during proton testing for both the substrate-removed detector array and the array with 30 μμm substrate. The detector array with full 800 μμm substrate exhibited substantial photocurrent for a flux of 103 protons/cm2^2-s at a beam energy of 18.1 MeV (~ 750 e−^-/s) and 34.4 MeV (∼\sim 65 e−^-/s). For the integrated space-like ambient proton flux level measured by the Spitzer Space Telescope, the luminescence would be well below the NEOCam dark current requirement of <200 e−^-/s, but the pattern of luminescence could be problematic, possibly complicating calibration.