Abstract

Purpose: This work proposes a novel RF pulse design for parallel transmit (pTx) systems to obtain uniform saturation of semisolid magnetization for Magnetization Transfer (MT) contrast in the presence of transmit field (B1+B_1^+) inhomogeneities. The semisolid magnetization is usually modeled as being purely longitudinal, with the applied B1+B_1^+ field saturating but not rotating its magnetization, thus standard pTx pulse design methods do not apply. Theory and Methods: Pulse design for Saturation Homogeneity (PUSH) optimizes pTx RF pulses by considering uniformity of root-mean squared B1+B_1^+, B1rmsB_1^{rms}, which relates to the rate of semisolid saturation. Here we considered designs consisting of a small number of spatially non-selective sub-pulses optimized over either a single 2D plane or 3D. Simulations and in vivo experiments on a 7T Terra system with an 8-TX Nova head coil in 5 subjects were carried out to study the homogenization of B1rmsB_1^{rms} and of the MT contrast by acquiring MT ratio maps. Results: Simulations and in vivo experiments showed up to 6 and 2 times more uniform B1rmsB_1^{rms} compared to circular polarized (CP) mode for 2D and 3D optimizations, respectively. This translated into 4 and 1.25 times more uniform MT contrast, consistently for all subjects, where 2 sub-pulses were enough for the implementation and coil used. Conclusion: The proposed PUSH method obtains more uniform and higher MT contrast than CP mode within the same SAR budget.

Explore this paper’s authors, topics and related work.

Building this map preview

BZPEER is loading the nearby papers, people, topics and institutions for this page.

Reviews 0

Write a reviewWrite

No reviews yet.

Discussion 0

Add a commentWrite

No comments yet.