Fast and Efficient Detection of 511 keV Photons using Cherenkov Light in PbF2_2 Crystal coupled to a MCP-PMT and SAMPIC Digitization Module

preprint2019arXivOpen access

Abstract

We study the possibility to use the Cherenkov light for the efficient detection of 511 keV photons with the goal to use it in TOF-PET. We designed and tested two detection modules consisting of PbF2_2 crystals attached to Planacon MCP-PMT XP85012. Amplified PMT signals are digitized by the SAMPIC module with high readout rate, up to 10510^5 events/s, and a negligible contribution to the time resolution, below 20 ps (FWHM). We developed a fast 2D scanning system to calibrate the PMT time response and studied in details the timing characteristics of the Planacon PMT. Using a radioactive 22^{22}Na source we measured a detection efficiency of 24% for 511 keV photons in a 10 mm thick crystal and a coincidence resolving time of 280 ps. We analyzed the main factors limiting the time resolution of the large-surface detection module and proposed solutions to improve it, which will be tested in our future project.

Explore this paper’s authors, topics and related work.

Building this map preview

BZPEER is loading the nearby papers, people, topics and institutions for this page.

Reviews 0

Write a reviewWrite

No reviews yet.

Discussion 0

Add a commentWrite

No comments yet.