Abstract

We have constructed an apparatus to study DC electrical breakdown in liquid helium at temperatures as low as 0.4 K and at pressures between the saturated vapor pressure and ∼\sim600 torr. The apparatus can house a set of electrodes that are 12 cm in diameter with a gap of 1−21-2 cm between them, and a potential up to ±50\pm 50 kV can be applied to each electrode. Initial results demonstrated that it is possible to apply fields exceeding 100 kV/cm in a 1 cm gap between two electropolished stainless steel electrodes 12 cm in diameter for a wide range of pressures at 0.4 K. We also measured the current between two electrodes. Our initial results, I<1I<1 pA at 45 kV, correspond to a lower bound on the effective volume resistivity of LHe of ρV>5×1018ρ_V > 5\times10^{18} Ω⋅Ω\cdotcm. This lower bound is 5 times larger than the bound previously measured. We report the design, construction, and operational experience of the apparatus, as well as initial results.