Abstract

A time-of-flight system with a plastic scintillator coupled to photomultipliers is developed for the external target facility (ETF). This system can satisfy the requirement of an ultrahigh vacuum (∼10−9\sim 10^{-9} mbar), a high counting rate (∼106\sim 10^{6} particles per second) and a magnetic field environment. In the beam test experiment, a total time resolution of 580 ps FWHM was obtained for the whole system, and nuclei with a mass of up to 80 could be identified using this system.