Abstract
In the work presented here we built and evaluated a single-channel neutron detection unit consisting of a ZnS:LiF scintillator with embedded WLS fibers readout by a SiPM. The unit has a sensitive volume of 2.4 x 2.8 x 50 mm; 12 WLS fibers of diameter 0.25 mm are uniformly distributed over this volume and are coupled to a 1 x 1 mm active area SiPM. We report the following performance parameters: neutron detection efficiency % at , background count rate Hz, gamma-sensitivity with Co source , dead time s, multi-count ratio %. All these parameters were achieved up to the SiPM dark count rate of MHz. We consider such detection unit as an elementary building block for realization of one-dimensional multichannel detectors for applications in the neutron scattering experimental technique. The dimensions of the unit and the number of embedded fibers can be varied to meet the specific application requirements. The upper limit of MHz on the SiPM dark count rate allows to use SiPMs with larger active areas if required.
Connections
Explore this paper’s authors, topics and related work.
Building this map preview
BZPEER is loading the nearby papers, people, topics and institutions for this page.
Reviews 0
Write a review
No reviews yet.
Discussion 0
Add a comment
No comments yet.