Comment on 'Field ion microscopy characterized tips in noncontact atomic force microscopy: Quantification of long-range force interactions'

preprint2013arXivOpen access

Abstract

A recent article by Falter et al. (Phys. Rev. B 87, 115412 (2013)) presents experimental results using field ion microscopy characterized tips in noncontact atomic force microscopy in order to characterize electrostatic and van der Waals long range forces. In the article, the tip radius was substantially underestimated at ~4.7 nm rather than ~8.1 nm due to subtleties in the application of the ring counting method. We point out where common errors in ring counting arise in order to benefit future experimental work in which the determination of tip radius by FIM is important.